Measure of blood pressure in function of sizes of cuffs, in patients in mediate perioperative phase, reception area of operation room and posoperative phase

Authors

  • Aparecida de Cassia Giani Peniche Universidade de São Paulo; Escola de Enfermagem
  • Edna Apparecida Moura Arcuri Universidade de São Paulo; Escola de Enfermagem; Enfermagem Médico-Cirúrgica

DOI:

https://doi.org/10.1590/0080-6234199202600200243

Keywords:

Blood pressure, Blood pressure determination

Abstract

The goal of this study was to compare blood pressure measures using two sizes of cuffs: one standard width (12 cm) and other with correct width that is 0,38 of arm circumference as recommended by American Heart Association. The comparisons were done among surgical patients in mediate perioperative phase, reception area of operation room and postoperative phase during the staying of the patient in the recovery room. The results demonstrated that the arterial blood pressure was hipoestimated by the use of the standard width cuff, reaching values up 30mmHg in the systolic arterial pressure and 30 mmHg in the diastolic ones.

Downloads

Download data is not yet available.

Published

1992-08-01

Issue

Section

Original Articles

How to Cite

Peniche, A. de C. G., & Arcuri, E. A. M. (1992). Measure of blood pressure in function of sizes of cuffs, in patients in mediate perioperative phase, reception area of operation room and posoperative phase. Revista Da Escola De Enfermagem Da USP, 26(2), 243-256. https://doi.org/10.1590/0080-6234199202600200243