Sterility Maintenance Assessment of Moist/Wet Material After Steam Sterilization and 30-day Storage
DOI:
https://doi.org/10.1590/S0104-11692010000400018Keywords:
Sterilization, Surgical Instruments, Contamination, Cross InfectionAbstract
Moist/wet materials stored after autoclaving are considered contaminated and not recommended for use. This study evaluates the maintenance of sterility in moist/wet material after being submitted to steam sterilization and stored for a period of 30 days. Aiming to support decision-making in emergency situations, 40 surgical boxes packed in nonwoven cloth covering Spunbound, Metblouwn, Spunbound (SMS): half (the experimental group) were placed in an autoclave but the drying phase was interrupted, yielding moist/wet materials and the other half (the negative control group) underwent the complete cycle. The external parts of each surgical box were deliberately contaminated with Serratia marcescens and subsequently stored for 30 days. After this period, the boxes' contents were submitted to sterility tests and no growth was observed. The presence of moisture inside the surgical boxes did not interfere with maintaining their sterility.Downloads
Download data is not yet available.
Downloads
Published
2010-08-01
Issue
Section
Original Articles
License
RLAE’s authorship concept is based on the substantial contribution by each of the individuals listed as authors, mainly in terms of conceiving and planning the research project, collecting or analyzing and interpreting data, writing and critical review. Indication of authors’ names under the article title is limited to six. If more, authors are listed on the online submission form under Acknowledgements. The possibility of including more than six authors will only be examined on multicenter studies, considering the explanations presented by the authors.Including names of authors whose contribution does not fit into the above criteria cannot be justified. Those names can be included in the Acknowledgements section.
Authors are fully responsible for the concepts disseminated in their manuscripts, which do not necessarily reflect the editors’ and editorial board’s opinion.
How to Cite
Sterility Maintenance Assessment of Moist/Wet Material After Steam Sterilization and 30-day Storage . (2010). Revista Latino-Americana De Enfermagem, 18(4), 786-791. https://doi.org/10.1590/S0104-11692010000400018